Z-2440-2012 Class II Terminated

Recalled by OPTI Medical Systems, Inc — Roswell, GA

FDA device recall Z-2440-2012 was initiated by OPTI Medical Systems, Inc on Jul 3, 2012 and is designated Class II. Reason for recall: The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters). The recall status is terminated (terminated Apr 11, 2013). Affected quantity: 895 boxes (25 cassettes per box).

Recall details

Product Type
Devices
Report Date
Oct 3, 2012
Initiation Date
Jul 3, 2012
Termination Date
Apr 11, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
895 boxes (25 cassettes per box)

Product description

BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to be used for in vitro measurements of pH, PCOX2 PO2, lactate (lactic acid), total hemoglobin (tHb), and oxygen saturation (SO2), in heparinized whole blood samples (either arterial or venous) on the OPTI CCA-TS system, in either a clinical setting or point of care locations. Cassettes contain one-time use sensors that are used with the OPTI CCA-TS analyzer to perform in-vitro measurements of blood gases and metabolites. OPTI CCA-TS B-Lac Cassettes are supplied in boxes of 25 and may only used with the OPTI CCA-TS Analyzer.

Reason for recall

The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters).

Distribution pattern

Worldwide Distribution -- US (nationwide) including the states of AZ, ID, TX and UT., and the countries of Argentina, Bolivia, China, Ecuador, France, Germany Greece, India, Indonesia, Iran, Ireland, Japan, Kuwait, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Nigeria, Poland, Romania, Russia, Saudi Arabia, Singapore, South Africa, Sudan Switzerland, Turkey, United kingdom, Venezuela, and Viet Nam.

Code information

Lot numbers: 110650, 114650, 114651, 114652, 114653, 116650, 116652, 117650, 118651, 121650, 121651, 122652, 122652, 122653, 122654, 122655, 122656, 122657, 123654, 123655, 124650, 133650, 134651, 136650, 146650, 149650, 202650, 208650, 208651, 209560, 209651, 211650, 212651, 213650, 214650, 215650, 215651, 221650, 222650 and 224650.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2440-2012
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for OPTI Medical Systems, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for OPTI Medical Systems, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about OPTI Medical Systems, Inc →

Source: openFDA record for Z-2440-2012. Report a data issue: hello@regdatalab.com.