Z-2445-2019 Class II Terminated
FDA device recall Z-2445-2019 was initiated by Roche Diagnostics Operations, Inc. on Jul 2, 2019 and is designated Class II. Reason for recall: The firm conducted internal investigations that confirmed customer complaints of quality control (QC) imprecision and calibration failures for the Tina-quant Myoglobin Gen.2 assay (catalog number… The recall status is terminated (terminated Oct 22, 2021). Affected quantity: 4,309 (cobas c311-340 units;cobas c501-3,517 units;cobas c502 452 units.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 11, 2019
- Initiation Date
- Jul 2, 2019
- Termination Date
- Oct 22, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,309 (cobas c311-340 units;cobas c501-3,517 units;cobas c502 452 units
Product Description
The Tina-Quant Myoglobin Gen. 2 Tests system is an immuno-turbidimetric assay for the quantitative in vitro determination of myoglobin in human serum and plasma on Roche automated clinical chemistry analyzers
Reason for Recall
The firm conducted internal investigations that confirmed customer complaints of quality control (QC) imprecision and calibration failures for the Tina-quant Myoglobin Gen.2 assay (catalog number 04580010190), lot number 349860 on the cobas c311 analyzer and cobas c501 and c502 modules due to a drop in signal in the reaction kinetics. This signal drop was traced to an adjacent cell mixing effect. Sample recovery is decreased as a consequence of the disturbance in reaction kinetics. The investigation found discrepancies of up to -30%. Calibration, QC, and patient samples can be affected. The following issues were reported: - Calibration failures; Dup.E error due to signal drop in reaction kinetics - Sporadic quality control (QC) imprecision. As a long-term solution, the firm will implement required extra wash cycles (EWCs) into the respective Special Wash Requirements method sheets for the cobas c 311 analyzer and cobas c 501 and 502 modules, and into the cobas e-file for the cobas c 502 module. For the use of the reagent with the cobas c501 and c502, Roche will also direct users to run the Tina-Quant Myoglobin gen. 2 assay independently from other tests, in a batch mode status, with a maximum of 39 samples per run. Additionally, a change in the application settings of the Tina-quant Myoglobin Gen.2 assay on the cobas c 501 and 502 modules will be implemented so that adjacent cuvettes are no longer used in order to prevent adjacent cell mixing effects. The application settings change will be communicated at a later date.
Distribution Pattern
Nationwide
Code Information
Lot: 349860 and onward