Z-2445-2021 Class II Ongoing
Recalled by Siemens Medical Solutions USA, Inc
— Malvern, PA
FDA device recall Z-2445-2021 was initiated by Siemens Medical Solutions USA, Inc on Aug 16, 2021 and is designated Class II. Reason for recall: Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the… The recall status is ongoing. Affected quantity: 126 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 22, 2021
- Initiation Date
- Aug 16, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 126 units
Product description
syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180
Reason for recall
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Distribution pattern
US Nationwide distribution.
Code information
Serial Numbers: 100713 223736 222946 223440 223032 223347 223084 223811 130643 223004 451078 101724 132011 131247 451036 132231 451032 222835 130848 500010 101773 132729 130874 133176 100421 100419 100420 100643 100642 100811 100814 100644 102369 100812 100641 100418 100813 100645 131284 130712 223734 102305 132893 223310 222411 223258 223334 223090 132781 131867 222563 222417 222491 131503 130125 451070 222224 132108 500013 132061 131169 451023 451017 451077 132701 102113 451094 130628 500007 131213 131360 500020 500028 451079 132205 451031 451001 222914 130331 451034 133191 102108 132136 131325 222686 223207 130734 102021 131578 500005 451071 222521 132904 102031 222539 222860 451039 130335 222862 222638 100677 222729 131336 223188 130452 223220 131801 131589 222719 132102 130140 223844 500004 451074 451030 133198 222601 133261 132595 130751 223150 131966 223424 223259 223058 131012 ***Updated 12/9/21*** 223021 ***Updated 1/31/22: Serial number 222475 133503
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2445-2021
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Siemens Medical Solutions USA, Inc →
Use it in ChatGPT
Get a free API key
How the connector works