Z-2447-2015 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-2447-2015 was initiated by Roche Diagnostics Operations, Inc. on Jul 14, 2015 and is designated Class II. Reason for recall: Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be… The recall status is terminated (terminated Jan 19, 2016). Affected quantity: 82 Fluid Packs.

Recall Details

Product Type
Devices
Report Date
Sep 2, 2015
Initiation Date
Jul 14, 2015
Termination Date
Jan 19, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
82 Fluid Packs

Product Description

cobas b 123 Fluid Pack COOX REF 05170036001 400 Product Usage: The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes NAT, K+, iCaWt (ISE), hematocrit (THct), metabolites (GLu, Lac) total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). The cobas b 123 POC system calculates derived parameters. Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC materials.

Reason for Recall

Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg

Distribution Pattern

US Nationwide Distribution in the states of LA, SC, CA, TX, NE, VA and WV including: Puerto Rico.

Code Information

cobas b 123 Fluid Pack COOx Catalog 05170036001, lot numbers 21446113 (exp 9/5/2015), and 21446153( exp 9/24/2015)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2447-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.