Z-2447-2021 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
September 22, 2021
Initiation Date
August 16, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 systems

Product Description

MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815

Reason for Recall

Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).

Distribution Pattern

US Nationwide distribution.

Code Information

Serial Numbers: 176302 175681