Z-2452-2015 Class II Terminated
FDA device recall Z-2452-2015 was initiated by Trilliant Surgical Ltd. on Jul 20, 2015 and is designated Class II. Reason for recall: The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through. The recall status is terminated (terminated Jan 26, 2017). Affected quantity: 121 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 2, 2015
- Initiation Date
- Jul 20, 2015
- Termination Date
- Jan 26, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 121 units
Product description
Reason for recall
Distribution pattern
Nationwide Distribution including NY, TX, OH, IN, FL, NH, MN, SC, OK, NJ, CA, NC, MO, GA, WI, IA, TN, WA, and Puerto Rico.