Z-2457-2012 Class II Terminated

Recalled by Integra Limited — Plainsboro, NJ

FDA device recall Z-2457-2012 was initiated by Integra Limited on Jul 27, 2012 and is designated Class II. Reason for recall: An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a… The recall status is terminated (terminated Feb 20, 2013). Affected quantity: 11 from Lot 0938990.

Recall details

Product Type
Devices
Report Date
Oct 3, 2012
Initiation Date
Jul 27, 2012
Termination Date
Feb 20, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11 from Lot 0938990

Product description

The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. Those devices provide cranial stability and immobility for various craniotomies and spinal procedures. The Gel Pads are supplied to users in labeled plastic bags, 1 per bag. Product Usage: The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. They are plastic gel filled devices that provide cushioning between the head and other metallic part of the MAYFIELD¿ system. The Left and Right Pads are mirror images.

Reason for recall

An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl

Distribution pattern

Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DC, GA, IA, IL, IN, NC, and NY and the countries of: Canada and Belgium.

Code information

The Lot No's for the affected Gel Pads are Lot 093990 of part number #440C1012 for Left Gel Pads

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2457-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2457-2012. Report a data issue: hello@regdatalab.com.