Z-2463-2012 Class II Terminated

Recalled by Imaging Sciences International, LLC — Hatfield, PA

FDA device recall Z-2463-2012 was initiated by Imaging Sciences International, LLC on Jul 30, 2012 and is designated Class II. Reason for recall: A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name… The recall status is terminated (terminated Oct 13, 2016). Affected quantity: 366.

Recall details

Product Type
Devices
Report Date
Oct 3, 2012
Initiation Date
Jul 30, 2012
Termination Date
Oct 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
366

Product description

DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System

Reason for recall

A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)

Distribution pattern

Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.

Code information

Catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2463-2012
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.