Z-2463-2012 Class II Terminated

Recalled by Imaging Sciences International, LLC — Hatfield, PA

Recall Details

Product Type
Devices
Report Date
October 3, 2012
Initiation Date
July 30, 2012
Termination Date
October 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
366

Product Description

DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System

Reason for Recall

A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)

Distribution Pattern

Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.

Code Information

Catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3