Z-2470-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 6, 2023
- Initiation Date
- August 1, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 24 units
Product Description
GS70H Salus Surgical Table, Product Code 18-071-70
Reason for Recall
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.
Code Information
Serial Numbers:SUC-20207-0002 SUC-20208-0006 SUC-20209-0010 SUC-2020X-0014 SUC-2022Z-0018 SUC-20233-0022 SUC-20207-0003 SUC-20208-0007 SUC-20209-0011 SUC-2020X-0015 SUC-2022Z-0019 SUC-20233-0023 SUC-20208-0004 SUC-20209-0008 SUC-2020X-0012 SUC-2020X-0016 SUC-2022Z-0020 SUC-20233-0024 SUC-20208-0005 SUC-20209-0009 SUC-2020X-0013 SUC-2022Z-0017 SUC-2022Z-0021 SUC-20233-0025