Z-2470-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 3, 2025
- Initiation Date
- July 14, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1094 units
Product Description
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.
Reason for Recall
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Distribution Pattern
US distribution to states of: IL and LA
Code Information
1) MODEL/ITEM NUMBER EVLP61I: UDI/DI 00191072194831, Kit Lot Numbers: 12-7359211, 12-7077211 , 12-7115211, 12-7162811, 12-7224111, 42-8276811, 42-8200111; 2) MODEL/ITEM NUMBER EVPV29G: UDI/DI 00191072194930, Kit Lot Numbers: 12-7288711, 12-7288711