Z-2472-2020 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

Recall Details

Product Type
Devices
Report Date
July 8, 2020
Initiation Date
May 8, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

uCT 530 Computed Tomography X-Ray System

Reason for Recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Distribution Pattern

Distributed to Texas.

Code Information

Units with software version no. R001.11.0.729741-Re-1100 including the following: Serial number/GTIN: 300162 (GTIN: 6971576831012); 300163 (GTIN: 6971576831012); 300174 (GTIN: 6971576831012); and 300176 (GTIN: 6971576831012).