Z-2472-2025 Class II Ongoing

Recalled by American Contract Systems Inc — Tiffin, OH

FDA device recall Z-2472-2025 was initiated by American Contract Systems Inc on Jul 14, 2025 and is designated Class II. Reason for recall: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization… The recall status is ongoing. Affected quantity: 588 units.

Recall details

Product Type
Devices
Report Date
Sep 3, 2025
Initiation Date
Jul 14, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
588 units

Product description

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H.

Reason for recall

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Distribution pattern

US distribution to states of: IL and LA

Code information

1) MODEL/ITEM NUMBER AMAH40N: UDI/DI 00191072235473, Kit Lot Numbers: 42-7886011, 42-7990911; 2)MODEL/ITEM NUMBER CCKM67F: UDI/DI 00191072233578, Kit Lot Numbers: 70-053862, 70-053457, 70-053042, 70-052715, 70-052358, 70-051850, 70-051243; 3)MODEL/ITEM NUMBER HSTH03N: UDI/DI 00191072189615, Kit Lot Numbers: 42-7115711; 4)MODEL/ITEM NUMBER MIAH65B: UDI/DI 00191072235411, Kit Lot Numbers: 42-8134311; 5)MODEL/ITEM NUMBER MPWH01B: UDI/DI 00191072233271, Kit Lot Numbers: 5436801; 6)MODEL/ITEM NUMBER SPKA57T: UDI/DI 00191072235848, Kit Lot Numbers: 12-7400011, 12-7381311, 12-7358211, 12-7082111, 12-7157111, 12-7309911; 7)MODEL/ITEM NUMBER WVAH40H: UDI/DI 00191072233004, Kit Lot Numbers: 42-7897411, 42-7966211.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2472-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2472-2025. Report a data issue: hello@regdatalab.com.