Z-2475-2021 Class II Terminated

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-2475-2021 was initiated by Philips North America Llc on Aug 13, 2021 and is designated Class II. Reason for recall: Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs… The recall status is terminated (terminated Mar 11, 2024). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Sep 22, 2021
Initiation Date
Aug 13, 2021
Termination Date
Mar 11, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis." "Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures."

Reason for recall

Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.

Distribution pattern

US Nationwide distribution in the states of ID, MA, MI, TX, VA.

Code information

Serial number: 7

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2475-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America Llc →

Source: openFDA record for Z-2475-2021. Report a data issue: hello@regdatalab.com.