Z-2475-2023 Class II Ongoing
FDA device recall Z-2475-2023 was initiated by Alphatec Spine, Inc. on Jul 7, 2023 and is designated Class II. Reason for recall: Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to… The recall status is ongoing. Affected quantity: 29.
Recall details
- Product Type
- Devices
- Report Date
- Sep 6, 2023
- Initiation Date
- Jul 7, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 29
Product description
LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,
Reason for recall
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.
Distribution pattern
US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.
Code information
UDI-DI: 00190376228037, Lot: EM50715
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2475-2023
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab