Z-2475-2023 Class II Ongoing

Recalled by Alphatec Spine, Inc. — Carlsbad, CA

FDA device recall Z-2475-2023 was initiated by Alphatec Spine, Inc. on Jul 7, 2023 and is designated Class II. Reason for recall: Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to… The recall status is ongoing. Affected quantity: 29.

Recall details

Product Type
Devices
Report Date
Sep 6, 2023
Initiation Date
Jul 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29

Product description

LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,

Reason for recall

Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.

Distribution pattern

US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.

Code information

UDI-DI: 00190376228037, Lot: EM50715

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2475-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2475-2023. Report a data issue: hello@regdatalab.com.