Z-2476-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 6, 2023
- Initiation Date
- June 1, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3201 ea
Product Description
epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Reason for Recall
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Distribution Pattern
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam