Z-2481-2023 Class II Ongoing

Recalled by Waldemar Link GmbH & Co. KG (Mfg Site) — Norderstedt, N/A

Recall Details

Product Type
Devices
Report Date
September 6, 2023
Initiation Date
July 26, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01

Reason for Recall

A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery

Distribution Pattern

IN,GA, TX

Code Information

UDI-DI: 04026575284658 Lot Codes: C010108, C151519