Z-2487-2012 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-2487-2012 was initiated by Stryker Howmedica Osteonics Corp. on Apr 2, 2012 and is designated Class II. Reason for recall: Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and… The recall status is terminated (terminated Jun 13, 2013). Affected quantity: 10, 983 units.

Recall details

Product Type
Devices
Report Date
Oct 10, 2012
Initiation Date
Apr 2, 2012
Termination Date
Jun 13, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10, 983 units

Product description

Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland. The Cone Body Trials simulate the respective Cone Body implants and mate with the stem trials to assist the surgeon in assessing the fit of the proposed Implants. The cone body trials is an assembly of the Body and 8 mm Hex Locking Bolt. The trial stem is attached to the trial Cone body via the distal thread of the Locking bolt.

Reason for recall

Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.

Distribution pattern

Nationwide distribution

Code information

Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm std cone trial 6278-1-025 Mod Rev 25 mm std cone trial 6278-1-027 Mod Rev 27 mm std cone trial 6278-1-029 Mod Rev 29 mm std cone trial 6278-1-031 Mod Rev 31 mm std cone trial 6278-1-119 Mod Rev 19 mm + 10 mm cone trial 6278-1-121 Mod Rev 21mm + 10 mm cone trial 6278-1-123 Mod Rev 23mm + 10 mm cone trial 6278-1 -125 Mod Rev 25 mm + 10 mm cone trial 6278-1-127 Mod Rev 27mm + 10 mm cone trial 6278-1-129 Mod Rev 29mm + 10mm cone trial 6278-1-131 Mod Rev 31mm + 10 mm cone trial 6278 -1-219 Mod Rev 19mm + 20 mm cone trial 6278-1-221 Mod Rev 21mm +20 mm cone trial 6278-1-223 Mod Rev 23mm + 20 mm cone trial 6278-1-225 Mod Rev 25mm + 20 mm cone trial 6278-1-227 Mod Rev 27mm + 20 mm cone trial 6278-1-229 Mod Rev 29mm + 20 mm cone trial 6278-1-231 Mod Rev 31mm + 20 mm cone trial 6278-1-319 Mod Rev 19mm + 30 mm cone trial 6278-1-321 Mod Rev 21mm +30 mm cone trial 6278-1-323 Mod Rev 23mm + 30 mm cone trial 6278-1-325 Mod Rev 25mm + 30 mm cone trial 6278-1-327 Mod Rev 27mm + 30 mm cone trial 6278-1-329 Mod rev 29mm +30 mm cone trial 6278-1-331 Mod Rev 31mm + 30 mm cone trial 1 single unit to each non-sterile package Numerous lots numbers please see attachment.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2487-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2487-2012. Report a data issue: hello@regdatalab.com.