Z-2500-2021 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2500-2021 was initiated by Siemens Medical Solutions USA, Inc on Aug 18, 2021 and is designated Class II. Reason for recall: Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows… The recall status is ongoing. Affected quantity: 31 units US.

Recall details

Product Type
Devices
Report Date
Sep 29, 2021
Initiation Date
Aug 18, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31 units US

Product description

Computed tomography x-ray systems with software syngo.CT VA20A_SP4a, VA20A_SP5, VA30A_SP2, VA30A_SP2a, VA30A_SP3,VA30A_FP2 in Somatom systems: SOMATOM go.Up - Model 11061620 SOMATOM go.Up - Model 11061628 SOMATOM go.All - Model 11061630 SOMATOM go.Top - Model 11061640 SOMATOM X.cite - Model 11330001 Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Reason for recall

Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows, resulting in delay in diagnosis or scan aborts with the necessity for patient rescan may occur

Distribution pattern

US Nationwide distribution.

Code information

US Serial Numbers: 111307, 111526, 119328, 108158, 123104, 111219, 123054, 117633, 111566, 111427, 123059, 117558, 117533, 119317, 117647, 108088, 123060, 123118, 123055, 111518, 123113, 108108, 123129, 123122, 123069, 111507, 111560, 119142, 108111, 108062, 119376. Update CT016/21/S is a remote update for the following products with software versions syngo CT VA30A SP2 or syngo CT VA30A SP2a or syngo CT VA30A SP3: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648), SOMATOM go.Sim (Model #11061660, #11061668), SOMATOM go.Open Pro (Model #11061670, #11061678). Update CT018/21/S is a remote update for the following products with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT022/21/S is an onsite update for the following product with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT025/21/S is an onsite update for the following product with software versions syngo.CT VA30A_SP2, VA30A_SP2a, syngo CT VA30A_SP3, or syngo CT VA30A_FP2 : SOMATOM X.cite (Model #11330001)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2500-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-2500-2021. Report a data issue: hello@regdatalab.com.