Z-2500-2021 Class II Ongoing
Recalled by Siemens Medical Solutions USA, Inc
— Malvern, PA
FDA device recall Z-2500-2021 was initiated by Siemens Medical Solutions USA, Inc on Aug 18, 2021 and is designated Class II. Reason for recall: Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows… The recall status is ongoing. Affected quantity: 31 units US.
Recall details
- Product Type
- Devices
- Report Date
- Sep 29, 2021
- Initiation Date
- Aug 18, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 31 units US
Product description
Computed tomography x-ray systems with software syngo.CT VA20A_SP4a, VA20A_SP5, VA30A_SP2, VA30A_SP2a, VA30A_SP3,VA30A_FP2 in Somatom systems: SOMATOM go.Up - Model 11061620 SOMATOM go.Up - Model 11061628 SOMATOM go.All - Model 11061630 SOMATOM go.Top - Model 11061640 SOMATOM X.cite - Model 11330001 Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
Reason for recall
Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows, resulting in delay in diagnosis or scan aborts with the necessity for patient rescan may occur
Distribution pattern
US Nationwide distribution.
Code information
US Serial Numbers: 111307, 111526, 119328, 108158, 123104, 111219, 123054, 117633, 111566, 111427, 123059, 117558, 117533, 119317, 117647, 108088, 123060, 123118, 123055, 111518, 123113, 108108, 123129, 123122, 123069, 111507, 111560, 119142, 108111, 108062, 119376. Update CT016/21/S is a remote update for the following products with software versions syngo CT VA30A SP2 or syngo CT VA30A SP2a or syngo CT VA30A SP3: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648), SOMATOM go.Sim (Model #11061660, #11061668), SOMATOM go.Open Pro (Model #11061670, #11061678). Update CT018/21/S is a remote update for the following products with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT022/21/S is an onsite update for the following product with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT025/21/S is an onsite update for the following product with software versions syngo.CT VA30A_SP2, VA30A_SP2a, syngo CT VA30A_SP3, or syngo CT VA30A_FP2 : SOMATOM X.cite (Model #11330001)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2500-2021
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Siemens Medical Solutions USA, Inc →
Use it in ChatGPT
Get a free API key
How the connector works