FDA device recall Z-2500-2026 was initiated by Medline Industries, LP on May 4, 2026 and is designated Class II. Reason for recall: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances… The recall status is ongoing. Affected quantity: 342 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 24, 2026
- Initiation Date
- May 4, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 342 units
Product description
MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Code information
Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24EBH848; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24DBK995; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23JBP023; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23HBR067; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23FBM330; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23EBK061; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22JBG042; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22GBH864; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22DBT571; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25DMJ438; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25CMH178; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24IME720; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24HMG912; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26CMC503; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26AMB047; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 25JME109.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2500-2026
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Medline Industries, LP? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works