Z-2505-2015 Class II Terminated

Recalled by Halyard Health, Inc — Alpharetta, GA

Recall Details

Product Type
Devices
Report Date
September 9, 2015
Initiation Date
July 29, 2015
Termination Date
January 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

KimVent* Wet Pak* Closed Suction System for Adults, 14 F, Elbow (Product Code 2276); KimVent* Wet Pak* Closed Suction System for Adults, 14 F, Elbow (Product Code 22169) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Reason for Recall

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Distribution Pattern

Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada

Code Information

Product Code 2276 -AB5131U06, AB5138U04, AB5161U12; Product Code 22169 - AB5131U09, AB5145U03