Z-2514-2025 Class II Ongoing

Recalled by Preat Corp — Santa Maria, CA

Recall Details

Product Type
Devices
Report Date
September 10, 2025
Initiation Date
June 25, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11

Product Description

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Reason for Recall

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

Distribution Pattern

US Nationwide distribution in the states of NY, FL, PA, CA, OH.

Code Information

UDI-DI: 00842092169599. Lot: 277097