Z-2514-2025 Class II Ongoing
FDA device recall Z-2514-2025 was initiated by Preat Corp on Jun 25, 2025 and is designated Class II. Reason for recall: Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will… The recall status is ongoing. Affected quantity: 11.
Recall details
- Product Type
- Devices
- Report Date
- Sep 10, 2025
- Initiation Date
- Jun 25, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of NY, FL, PA, CA, OH.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2514-2025
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Preat Corp? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.