Z-2521-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 9, 2015
- Initiation Date
- July 29, 2015
- Termination Date
- January 12, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
KimVent* Closed Suction System for Neonates/Pediatrics, 5 F, Manifold (Product Code 201) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Reason for Recall
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Distribution Pattern
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Code Information
Product Code 201 - M5076T405