Z-2521-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 1, 2018
- Initiation Date
- June 21, 2018
- Termination Date
- April 6, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
CORAIL Cementless Femoral Stem HA Coated Standard Size 11, Catalog No. 3L92511
Reason for Recall
Two lots of femoral stems were labeled as Size 11. However, they are actually Size 12. Potential adverse effects include poor joint mechanics, improper fit, loosening, and additional surgical procedure to correct. It may also result in an intraoperative fracture which could represent a surgical delay and additional procedures.
Distribution Pattern
Accounts in AL, CA, CO, GA< IL, LA, ME, NC, NH, NY, OR, TX, UT, and WA. Foreign distribution to Belgium, Croatia, Cyprus, Denmark, Germany, Indonesia, Korea, Sweden, and Tunisia.
Code Information
5307603 5300997