Z-2521-2020 Class II Terminated

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai

Recall Details

Product Type
Devices
Report Date
July 15, 2020
Initiation Date
February 27, 2020
Termination Date
May 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

uDR 596i Digital Medical X-Ray Imaging System - Product Usage: intended to use by a qualified/ trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk.

Reason for Recall

Two issues were identified with the stationary x-ray system. 1. Some position illustrations in the patient exam interface show the incorrect anatomical position for the standard of care. When the image is acquired with the wrong reference position, clinical diagnostic requirements cannot be met.T2. Where image stitching is used, the left and right tag information may be lost. When these problems are encountered, an additional exposure is needed and the patient will receive more radiation.

Distribution Pattern

US Nationwide distribution including in the state of Texas.

Code Information

Model Number uDR 596i with software versions R002.19.0-1.8402-690106 and R002.19.0.8648.690112. Serial numbers: 270012, 270013, 270014, 270015, 270016 and 270017. UDI numbers: (01)06971576833023(240)88000093(21)270012; (01)06971576833023(240)88000093(21)270013; (01)06971576833023(240)88000093(21)270014; (01)06971576833023(240)88000093(21)270015; XXX; and XXX.