Z-2533-2020 Class II Terminated

Recalled by TELEFLEX MEDICAL INC — Morrisville, NC

FDA device recall Z-2533-2020 was initiated by TELEFLEX MEDICAL INC on May 29, 2020 and is designated Class II. Reason for recall: Teleflex is initiating a voluntary recall for the above-mentioned products due to reports indicating that the laser guard foil partially separated and/or slightly detached at the edges in the presence… The recall status is terminated (terminated Jun 17, 2025). Affected quantity: 6014 devices.

Recall details

Product Type
Devices
Report Date
Jul 15, 2020
Initiation Date
May 29, 2020
Termination Date
Jun 17, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6014 devices

Product description

RUSCH LASERTUBE (Rubber), RES 102004, Sizes: a) I.D. mm 4,0, Product Code 102004-000040 b) I.D. mm 5,0, Product Code 102004-000050 c) I.D. mm 6,0, Product Code 102004-000060 d) I.D. mm 7,0, Product Code 102004-000070 e) I.D. mm 8,0, Product Code 102004-000080 Product Usage: Tracheal intubation during laser surgery in the laryngeal area. The tube is resistant to the following kinds of laser: A R+ laser / ND/YAG laser / CO2 lasers (continuous-wave or pulsed) with a wave length ranging from 0.488 ¿m to 10.6 ¿m.

Reason for recall

Teleflex is initiating a voluntary recall for the above-mentioned products due to reports indicating that the laser guard foil partially separated and/or slightly detached at the edges in the presence of moisture.

Distribution pattern

US Nationwide distribution.

Code information

a) I.D. mm 4,0, Product Code 102004-000040, Lot numbers: 19491, 19501, 19511, 20021, 20041, 20091 b) I.D. mm 5,0, Product Code 102004-000050, Lot numbers: 19471, 19501, 19511, 20021, 20031, 20041, 20051, 20061, 20071, 20101, 20111, 20121 c) I.D. mm 6,0, Product Code 102004-000060, Lot numbers: 19481, 19491, 19501, 20081, 20091, 20101, 20111, 20121, 20131 d) I.D. mm 7,0, Product Code 102004-000070, Lot numbers: 19451, 19461, 19471, 19501, 20021, 20031, 20051, 20091, 20111 e) I.D. mm 8,0, Product Code 102004-000080, Lot numbers: 19451, 19481, 19501, 19511, 20041, 20111, 20121

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2533-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for TELEFLEX MEDICAL INC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for TELEFLEX MEDICAL INC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about TELEFLEX MEDICAL INC →

Source: openFDA record for Z-2533-2020. Report a data issue: hello@regdatalab.com.