Z-2535-2020 Class II Terminated

Recalled by LABORIE MEDICAL TECHNOLOGIES, CANADA ULC — Mississauga

Recall Details

Product Type
Devices
Report Date
July 15, 2020
Initiation Date
April 22, 2020
Termination Date
April 25, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
480 units

Product Description

Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.

Reason for Recall

This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.

Distribution Pattern

US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.

Code Information

Model Number: REF 902101226, LOT 15883/1