Z-2535-2020 Class II Terminated

Recalled by LABORIE MEDICAL TECHNOLOGIES, CANADA ULC — Mississauga

FDA device recall Z-2535-2020 was initiated by LABORIE MEDICAL TECHNOLOGIES, CANADA ULC on Apr 22, 2020 and is designated Class II. Reason for recall: This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in… The recall status is terminated (terminated Apr 25, 2021). Affected quantity: 480 units.

Recall details

Product Type
Devices
Report Date
Jul 15, 2020
Initiation Date
Apr 22, 2020
Termination Date
Apr 25, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
480 units

Product description

Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.

Reason for recall

This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.

Distribution pattern

US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.

Code information

Model Number: REF 902101226, LOT 15883/1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2535-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LABORIE MEDICAL TECHNOLOGIES, CANADA ULC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for LABORIE MEDICAL TECHNOLOGIES, CANADA ULC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about LABORIE MEDICAL TECHNOLOGIES, CANADA ULC →

Source: openFDA record for Z-2535-2020. Report a data issue: hello@regdatalab.com.