Z-2536-2016 Class II Terminated
FDA device recall Z-2536-2016 was initiated by Stryker Instruments Div. Of Stryker Corporation on Jun 14, 2016 and is designated Class II. Reason for recall: The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if… The recall status is terminated (terminated Mar 27, 2017). Affected quantity: 4,841 boxes (9,682 units).
Recall details
- Product Type
- Devices
- Report Date
- Aug 24, 2016
- Initiation Date
- Jun 14, 2016
- Termination Date
- Mar 27, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,841 boxes (9,682 units)
Product description
Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.
Reason for recall
The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.
Distribution pattern
Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.
Code information
Model Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000 Lot numbers: 16021012, 16027022, 16049012, 16061012, 16092012, 16107012, 16022012, 16028012, 16049022, 16063012, 16098012, 16109012, 16025012, 16028022, 16049032, 16063022, 16100012, 16109022, 16025022, 16033012, 16053012, 16070012, 16104012, 16113022, 16026012, 16041012, 16056012, 16070022, 16104022, 16113032, 16026022, 16048012, 16056022, 16074012, 16105012, 16118012, 16027012, 16048022, 16060012, 16088012, 16106012, 16118022 0605-887-000 Lot numbers: 16041012, 16040012, 16050012, 16057012, 16069012, 16078012, 16104012, 16112012, 16124012 0605-687-000 Lot numbers: 16041012, 16047012, 16111012 0605-683-000 Lot numbers: 16071012
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2536-2016
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026