Z-2536-2025 Class II Ongoing

Recalled by Boston Scientific Corporation — Maple Grove, MN

Recall Details

Product Type
Devices
Report Date
September 17, 2025
Initiation Date
August 5, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1369 units

Product Description

Boston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260, 7. M0062251290, 8. M0062251300, 9. M0062251310. 10. M0062251350, 11. M0062251360, 12. M0062251370. (Urology)

Reason for Recall

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

Distribution Pattern

Worldwide distribution.

Code Information

1. M0062251200, GTIN 8714729341277, Lot Numbers: 35536688, 35538633, 35538634, 35543803, 35543804, 35572728, 35678100; 2. M0062251210, GTIN 8714729341284, Lot Numbers: 35274797, 35274799, 35277223, 35277224, 35294475, 35294476, 35584203, 35584215, 35656798, 35667482, 35667483; 3. M0062251220, GTIN 8714729341291, Lot Numbers: 35257438, 35535450, 35536202, 35547680, 35656799; 4. M0062251230, GTIN 8714729341307, Lot Numbers: 35266389, 35543805; 5. M0062251240, GTIN 8714729341314, Lot Numbers: 35690647; 6. M0062251260, GTIN 8714729341338, Lot Numbers: 35537838, 35537839. 7. M0062251290, GTIN 08714729341369, Lot Numbers: 35266391; 8. M0062251300, GTIN 08714729341376, Lot Numbers: 35275806, 35536992; 9. M0062251310, GTIN 08714729341383, Lot Numbers: 35536991; 10. M0062251350, GTIN 08714729341390, Lot Numbers: 35690648; 11. M0062251360, GTIN 08714729302421, Lot Numbers: 35257611, 35536990; 12. M0062251370, GTIN 08714729302438, Lot Numbers: 35294477, 35536689.