Z-2540-2019 Class II Ongoing
FDA device recall Z-2540-2019 was initiated by Stryker Communications on March 22, 2019 and is designated Class II. Reason for recall: The back cover of the light head was insufficiently assembled and could potentially fall off the light. The recall status is ongoing. Affected quantity: 3,344 lights.
Recall Details
- Product Type
- Devices
- Report Date
- September 18, 2019
- Initiation Date
- March 22, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,344 lights
Product Description
Reason for Recall
Distribution Pattern
Distribution was nationwide, including to Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Austria, Australia, Belgium, Bulgaria, Canada, China, Colombia, Denmark, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Kazakhstan, Korea, Kuwait, Lithuania, Mexico, Netherlands, Philippines, Poland, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.