Z-2551-2019 Class II Terminated

Recalled by Sentinel CH SpA — Milan

FDA device recall Z-2551-2019 was initiated by Sentinel CH SpA on Jun 28, 2019 and is designated Class II. Reason for recall: Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate… The recall status is terminated (terminated Jul 24, 2020). Affected quantity: 15919 kits.

Recall details

Product Type
Devices
Report Date
Oct 2, 2019
Initiation Date
Jun 28, 2019
Termination Date
Jul 24, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15919 kits

Product description

MULTIGENT Lithium-intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 8L25-30

Reason for recall

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

Distribution pattern

Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOURG MALAYSIA MALDIVES MEXICO MOROCCO NETHERLANDS NEW ZEALAND NIGERIA NORWAY OMAN PAKISTAN PANAMA PARAGUAY PHILIPPINES POLAND PORTUGAL PUERTO RICO ROMANIA RWANDA SAUDI ARABIA SLOVAKIA SOUTH AFRICA SPAIN ST. VINCENT SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UAE UNITED KINGDOM URUGUAY

Code information

Lot Number/UDI: 71175Y600 (01)003807400177712(17)191031(10)71175Y600; 80362Y600 (01)003807400177712(17)200131(10)80362Y600; 80201Y600 (01)003807400177712(17)200131(10)80201Y600; 80533Y600 (01)003807400177712(17)200630(10)80533Y600; 80463Y600 (01)003807400177712(17)200630(10)80463Y600; 80424Y600 (01)003807400177712(17)200630(10)80424Y600; 80605Y600 (01)003807400177712(17)200930(10)80605Y600; 90208Y600 (01)003807400177712(17)210131(10)90208Y600

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2551-2019
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2551-2019. Report a data issue: hello@regdatalab.com.