Z-2555-2014 Class II Terminated

Recalled by Thermedx LLC — Solon, OH

FDA device recall Z-2555-2014 was initiated by Thermedx LLC on Feb 10, 2014 and is designated Class II. Reason for recall: A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the… The recall status is terminated (terminated Aug 4, 2015). Affected quantity: 21.

Recall details

Product Type
Devices
Report Date
Sep 3, 2014
Initiation Date
Feb 10, 2014
Termination Date
Aug 4, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21

Product description

The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.

Reason for recall

A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.

Distribution pattern

Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.

Code information

Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2555-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Thermedx LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2555-2014. Report a data issue: hello@regdatalab.com.