Z-2569-2020 Class I Terminated

Recalled by CME America, LLC — Golden, CO

FDA device recall Z-2569-2020 was initiated by CME America, LLC on Jun 16, 2020 and is designated Class I. Reason for recall: Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of… The recall status is terminated (terminated May 9, 2022). Affected quantity: 25286.

Recall details

Product Type
Devices
Report Date
Jul 22, 2020
Initiation Date
Jun 16, 2020
Termination Date
May 9, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25286

Product description

CMExpress Microbore Set, REF: A120-161CYF

Reason for recall

Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy

Distribution pattern

U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK

Code information

All lots within expiry

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2569-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for CME America, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2569-2020. Report a data issue: hello@regdatalab.com.