Z-2584-2021 Class II Terminated

Recalled by Deerfield Imaging, Inc. — Minnetonka, MN

Recall Details

Product Type
Devices
Report Date
October 6, 2021
Initiation Date
August 25, 2021
Termination Date
April 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to support and position a patient.

Reason for Recall

ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

Distribution Pattern

Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.

Code Information

UDI: 00857534006738 and 00857534006349 Model Number: 113821-000 (Serial Number: 10002220); Model Number: 113821-600 (Serial Numbers: 10002692, 10006152, 10006991, 10004513, 10006997, and 10003754); Model Number: 114148-000 (Serial Numbers: 10004463, 10006891, 10005189, 10005190, 10004327, and 10004997)