Z-2588-2017 Class II Terminated
Recalled by MONICA HEALTHCARE LTD
— Nottingham, N/A
FDA device recall Z-2588-2017 was initiated by MONICA HEALTHCARE LTD on Apr 28, 2017 and is designated Class II. Reason for recall: Monica Novii Wireless Patch System-Potential battery damage and excess heat through prolonged charging. The recall status is terminated (terminated Mar 1, 2018). Affected quantity: 4,110 (4,106 US; 4 OUS).
Recall details
- Product Type
- Devices
- Report Date
- Jun 21, 2017
- Initiation Date
- Apr 28, 2017
- Termination Date
- Mar 1, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,110 (4,106 US; 4 OUS)
Product description
Monica Novii Wireless Patch System product Usage: Novii Interface - The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. Novii Pod - The Monica Novii Pod is an intrapartum maternal-fetal monitor that non-invasively measures and displays fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii Pod acquires and displays the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Novii Pod also acquires and displays the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Novii Pod is indicated for use on women who are at >36 completed weeks, in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen.
Reason for recall
Monica Novii Wireless Patch System-Potential battery damage and excess heat through prolonged charging.
Distribution pattern
The product was distributed thru the United States. Please see attached excel sheet for detailed information about the consignees.
Code information
See attached consignee list for affected device serial numbers.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2588-2017
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for MONICA HEALTHCARE LTD? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for MONICA HEALTHCARE LTD: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about MONICA HEALTHCARE LTD →
Use it in ChatGPT
Get a free API key
How the connector works