Z-2595-2025 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
October 1, 2025
Initiation Date
August 13, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
181 units

Product Description

Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317

Reason for Recall

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.

Code Information

1. Model Number: 728235; UDI-DI: N/A; Serial Numbers: 9223, 9115, 9225, 29144. 2. Model Number: 728231; UDI-DI: N/A; Serial Numbers: 95647, 9194, 95341, 95234, 10051, 95488, 10098, 95695, 10309, 95217, 95698, 10443, 10544, 10531, 10712, 10662, 95834, 9990, 9784, 95632, 9058, 9089, 9522, 9580, 9688, 9877, 9806, 9808, 9671, 9694, 9703, 9830, 29068, 9821, 9869, 9837, 90144, 95203, 95225, 95303, 95328, 95362, 95402, 95395, 9852, 95492, 95491, 9965, 9974, 95453, 95502, 10006, 10025, 95520, 10060, 95630, 95621, 10141, 95673, 10153, 10137, 95521, 90128, 9605, 10218, 10199, 10247, 10275, 10267, 10245, 10394, 10307, 90124, 10281, 95692, 29045, 95349, 95639, 10344, 10330, 10310, 10348, 10391, 10393, 10418, 10205, 9799, 10429, 29062, 10511, 10431, 10515, 10535, 10315, 10521, 10518, 10566, 10536, 10522, 10496, 10560, 10568, 90170, 10433, 10574, 10609, 10559, 10468, 95725, 10622, 95710, 95729, 10590, 10628, 10683, 10668, 10666, 10607, 95758, 10706, 95776, 95821, 95757, 95858, 95795, 9868, 10729, 95791, 10734, 10718, 95807, 10649, 95781, 95790, 95823, 95813, 10788, 10774, 95759, 95852, 95848, 95116, 95872, 95871, 95914, 95886, 95498, 95054, 95928, 95933, 95025, 90162, 95144, 29039, 95932, 95399, 10031, 90804, 10161, 10494, 9934, 90161. 3. Model Number: 728317; UDI-DI: N/A; Serial Numbers: 345040, 345053, 345155, 345031, 345048, 345056, 345055, 345063, 345050, 345100, 345117, 345122, 345133, 345131, 345171.