Z-2597-2016 Class II Terminated

Recalled by Conmed Corporation — Utica, NY

FDA device recall Z-2597-2016 was initiated by Conmed Corporation on May 18, 2016 and is designated Class II. Reason for recall: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn. The recall status is terminated (terminated May 24, 2018). Affected quantity: 6,640 units.

Recall details

Product Type
Devices
Report Date
Aug 31, 2016
Initiation Date
May 18, 2016
Termination Date
May 24, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6,640 units

Product description

Pediatric ThermoGard¿ Dual Dispersive Electrodes (for patients 5-15 kg.), 10' (3.05m) Cable, Catalog Number 51-7710. Intended to be used for the dispersion and return to the electrosurgical generator.

Reason for recall

A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.

Distribution pattern

Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam, China, France, Guatemala, Hong Kong, Israel, Italy, Japan, South Korea, Malaysia, Mexico, New Zealand, Norway, Portugal, Saudi Arabia, Spain, Sweden, and United Arab Emirates.

Code information

Beginning Lot Code 141001X with Ending Lot Code 20150705X.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2597-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Conmed Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2597-2016. Report a data issue: hello@regdatalab.com.