Z-2598-2025 Class II Ongoing
FDA device recall Z-2598-2025 was initiated by Philips North America Llc on Aug 13, 2025 and is designated Class II. Reason for recall: The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. The recall status is ongoing. Affected quantity: 210 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 1, 2025
- Initiation Date
- Aug 13, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 210 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2598-2025
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Philips North America Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Philips North America Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.