Z-2602-2023 Class II Ongoing

Recalled by Stryker Corporation — Portage, MI

Recall Details

Product Type
Devices
Report Date
September 27, 2023
Initiation Date
August 7, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
231 units US; 21 units OUS

Product Description

Stryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle, Restrictors & Prox. Med. Press- Allows the user to mix bone cement with a constant high vacuum through the process. Catalog Number: 0306-705-000

Reason for Recall

Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

Distribution Pattern

Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK

Code Information

GTIN: 04546540055422 Lot Numbers: 22306012 22307012 22340012 22341012 22342012 23017012