Z-2630-2016 Class II Terminated
FDA device recall Z-2630-2016 was initiated by Stryker Instruments Div. Of Stryker Corporation on Jul 1, 2016 and is designated Class II. Reason for recall: Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of… The recall status is terminated (terminated Jun 2, 2017). Affected quantity: 78.
Recall details
- Product Type
- Devices
- Report Date
- Aug 31, 2016
- Initiation Date
- Jul 1, 2016
- Termination Date
- Jun 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 78
Product description
120 V Neptune 3 Rover, Model Number: 0703-001-000
Reason for recall
Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.
Distribution pattern
Domestic:MI, CA, ID, IA VA/DOD: None Foreign: None
Code information
120 V Neptune 3 Rover, Model Number: 0703-001-000; Serial Numbers: 1606300523, 1610606263, 1610903973, 1610904293, 1612303683, 1612303823, 1612303963, 1606300553, 1610606273, 1610904063, 1610904313, 1612303703, 1612303833, 1612303973, 1606300583, 1610606303, 1610904083 , 1610904343, 1612303713, 1612303843 , 1612303993, 1606300593, 1610606323, 1610904103, 1610904353, 1612303723, 1612303853 , 1613106443, 1606300613, 1610606353, 1610904123, 1610904363 , 1612303733, 1612303863, 1613106463, 1606300623, 1610619473, 1610904163, 1610904373 , 1612303743, 1612303873, 1613106583, 1606300643, 1610619483, 1610904173, 1610904403, 1612303763 , 1612303883 , 1613201233, 1606300653, 1610619543, 1610904183, 1612303633, 1612303783, 1612303903, 1613201243 , 1607503463 , 1610619573 , 1610904193, 1612303643, 1612303793, 1612303913, 1613201303, 1607503493, 1610619603 , 1610904203, 1612303663, 1612303813, 612303933, 1613201323, 1610409473, 1610903893, 1610904213, 1612303673
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2630-2016
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026