Z-2641-2020 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

FDA device recall Z-2641-2020 was initiated by Shanghai United Imaging Healthcare Co., Ltd. on May 23, 2020 and is designated Class II. Reason for recall: Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent… The recall status is ongoing. Affected quantity: 3.

Recall details

Product Type
Devices
Report Date
Jul 29, 2020
Initiation Date
May 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product description

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Reason for recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Distribution pattern

US Distribution to state of: TX

Code information

Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2641-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Shanghai United Imaging Healthcare Co., Ltd.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Shanghai United Imaging Healthcare Co., Ltd.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Shanghai United Imaging Healthcare Co… →

Source: openFDA record for Z-2641-2020. Report a data issue: hello@regdatalab.com.