Z-2641-2020 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

Recall Details

Product Type
Devices
Report Date
July 29, 2020
Initiation Date
May 23, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Reason for Recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Distribution Pattern

US Distribution to state of: TX

Code Information

Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100