Z-2653-2014 Class II Terminated

Recalled by MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK — HIROSHIMA, N/A

Recall Details

Product Type
Devices
Report Date
September 24, 2014
Initiation Date
July 5, 2014
Termination Date
April 1, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units in use in the US

Product Description

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Reason for Recall

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

Distribution Pattern

Distributed in the states of NY, FL, and TX.

Code Information

Software Version 3.0.0 and after. Serial Numbers 201902, 203901, and 203919