Z-2655-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 11, 2023
- Initiation Date
- August 31, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13 total devices
Product Description
EarlyVue VS30 Vital Signs Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863359
Reason for Recall
A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.
Code Information
UDI-DI: 00884838075900; Serial Numbers: CN82203600