Z-2664-2017 Class II Terminated

Recalled by Exactech, Inc. — Gainesville, FL

Recall Details

Product Type
Devices
Report Date
July 12, 2017
Initiation Date
May 25, 2017
Termination Date
July 26, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
813

Product Description

EXACTECH Equinoxe Glenoid Reamer - Catalog #: 321-05-38, Catalog #: 321-05-42, and Catalog #: 321-05-46

Reason for Recall

Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.

Distribution Pattern

AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom

Code Information

All Lots