Z-2667-2020 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-2667-2020 was initiated by Becton Dickinson & Company on Jun 19, 2020 and is designated Class II. Reason for recall: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a… The recall status is terminated (terminated Jul 29, 2022). Affected quantity: 5,296,132 total.

Recall details

Product Type
Devices
Report Date
Jul 29, 2020
Initiation Date
Jun 19, 2020
Termination Date
Jul 29, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,296,132 total

Product description

SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDI 00801741143892)

Reason for recall

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Distribution pattern

Worldwide Distribution

Code information

Catalog Number CPA00040 (UDI 00801741143892). Lot Numbers: REEN0219 REEP1376 REEP3657

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2667-2020
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.