Z-2682-2016 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-2682-2016 was initiated by Philips Electronics North America Corporation on Sep 17, 2015 and is designated Class II. Reason for recall: Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the… The recall status is terminated (terminated Aug 31, 2017). Affected quantity: 15.

Recall details

Product Type
Devices
Report Date
Sep 7, 2016
Initiation Date
Sep 17, 2015
Termination Date
Aug 31, 2017
Voluntary/Mandated
FDA Mandated
Product Quantity
15

Product description

Philips DuraDiagnost stationary X-ray system

Reason for recall

Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].

Distribution pattern

US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.

Code information

SN: 140054 140059 150013 150014 150039 150041 150042 150052 150053 150055 150056 150059 150061 150091 150093

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2682-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2682-2016. Report a data issue: hello@regdatalab.com.