Z-2682-2016 Class II Terminated
FDA device recall Z-2682-2016 was initiated by Philips Electronics North America Corporation on Sep 17, 2015 and is designated Class II. Reason for recall: Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the… The recall status is terminated (terminated Aug 31, 2017). Affected quantity: 15.
Recall details
- Product Type
- Devices
- Report Date
- Sep 7, 2016
- Initiation Date
- Sep 17, 2015
- Termination Date
- Aug 31, 2017
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 15
Product description
Philips DuraDiagnost stationary X-ray system
Reason for recall
Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].
Distribution pattern
US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.
Code information
SN: 140054 140059 150013 150014 150039 150041 150042 150052 150053 150055 150056 150059 150061 150091 150093