Z-2687-2015 Class II Terminated
FDA device recall Z-2687-2015 was initiated by Customed, Inc on Jul 10, 2015 and is designated Class II. Reason for recall: Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile. The recall status is terminated (terminated Apr 12, 2017). Affected quantity: 912.
Recall details
- Product Type
- Devices
- Report Date
- Sep 9, 2015
- Initiation Date
- Jul 10, 2015
- Termination Date
- Apr 12, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 912
Product description
PERY-GYN SURGICAL PACK 6/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .
Reason for recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Distribution pattern
Distributed Only in Puerto Rico.
Code information
Code:900548A Lots: 141015718 exp. 11/30/15 141216350 exp. 12/31/15 150116506 exp. 12/31/15 150116945 exp. 1/31/16 150116946 exp. 2/28/16 150318003 exp. 3/31/16 150318001 exp. 3/31/16 150318002 exp. 3/31/16 150418416 exp. 4/30/16 150418417 exp. 4/30/16 150518829 exp. 5/31/16 150518829 exp. 5/31/16 150619185 exp. 6/30/16 150619404 exp. 6/30/16
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2687-2015
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Customed, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.