Z-2688-2020 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
July 29, 2020
Initiation Date
June 24, 2020
Termination Date
April 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 units

Product Description

syngo.via RT Image Suite with software versions syngo.via VB30 or VB40

Reason for Recall

If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.

Distribution Pattern

US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.

Code Information

syngo.via VB30 or VB40: All Lots Model #: 10496180