Z-2693-2014 Class II Terminated

Recalled by ARJOHUNTLEIGH POLSKA Sp. Z.O.O. — KOMORNIKI, N/A

FDA device recall Z-2693-2014 was initiated by ARJOHUNTLEIGH POLSKA Sp. Z.O.O. on Jul 28, 2014 and is designated Class II. Reason for recall: ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are… The recall status is terminated (terminated Sep 13, 2016). Affected quantity: 18 Lifts.

Recall details

Product Type
Devices
Report Date
Oct 1, 2014
Initiation Date
Jul 28, 2014
Termination Date
Sep 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18 Lifts

Product description

Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.

Reason for recall

ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp

Distribution pattern

Worldwide Distribution - US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.

Code information

Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2693-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2693-2014. Report a data issue: hello@regdatalab.com.