Z-2693-2014 Class II Terminated
FDA device recall Z-2693-2014 was initiated by ARJOHUNTLEIGH POLSKA Sp. Z.O.O. on Jul 28, 2014 and is designated Class II. Reason for recall: ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are… The recall status is terminated (terminated Sep 13, 2016). Affected quantity: 18 Lifts.
Recall details
- Product Type
- Devices
- Report Date
- Oct 1, 2014
- Initiation Date
- Jul 28, 2014
- Termination Date
- Sep 13, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 18 Lifts
Product description
Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.
Reason for recall
ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp
Distribution pattern
Worldwide Distribution - US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.
Code information
Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531