Z-2694-2020 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

Recall Details

Product Type
Devices
Report Date
July 29, 2020
Initiation Date
June 8, 2020
Termination Date
May 10, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 pk

Product Description

Accurian RF Ablation, Cannula pk, Accurian, 100mm, 22G, 5mm, CS Product Number: AC0016 GTIN / UPN: 20763000153674 - Product Usage: (sterile, single-use) comprises an electrically insulated shaft, an active electrode for RF energy delivery, a hub with a female slip luer fitting, and a stylet with a color-coded cap to indicate gauge size. The ACCURIAN RF Cannula included in the ACCURIAN RF Ablation Platform intended for the creation of radiofrequency lesions in nervous tissue.

Reason for Recall

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Belgium and Australia.

Code Information

Lot # BY19L039