Z-2695-2020 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

FDA device recall Z-2695-2020 was initiated by Medtronic Sofamor Danek USA Inc on Jun 8, 2020 and is designated Class II. Reason for recall: Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula. The recall status is terminated (terminated May 10, 2022). Affected quantity: 25 pk.

Recall details

Product Type
Devices
Report Date
Jul 29, 2020
Initiation Date
Jun 8, 2020
Termination Date
May 10, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 pk

Product description

Accurian RF Ablation,Cannula pk, Accurian, 100mm, 22G, 10mm, SS Product Number: AC0019 GTIN / UPN: 20763000153667 - Product Usage: (sterile, single-use) comprises an electrically insulated shaft, an active electrode for RF energy delivery, a hub with a female slip luer fitting, and a stylet with a color-coded cap to indicate gauge size. The ACCURIAN RF Cannula included in the ACCURIAN RF Ablation Platform intended for the creation of radiofrequency lesions in nervous tissue.

Reason for recall

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Distribution pattern

Worldwide distribution - US Nationwide distribution and the countries of Belgium and Australia.

Code information

Lot # BY19D020

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2695-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Sofamor Danek USA Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Sofamor Danek USA Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Sofamor Danek USA Inc →

Source: openFDA record for Z-2695-2020. Report a data issue: hello@regdatalab.com.